em-tec Markets

For more than 30 years,

em-tec has been at the forefront of flow measurement technology, developing state-of-the-art and solution-oriented flow meters for the medical and bioprocessing markets. Learn more about our markets and applications using the cards below.

View Our Markets

Learn More About the Accuracy Parameters for Each BioProTT™ Clamp-On Transducer

CalibrationAdjustment

1. The following parameters impact the ultrasonic signals inside the BioProTT™ Clamp-On Transducer:

  • Medium type
  • Medium temperature
  • Tubing material
  • Flow range

2. For best results, each sensor is individually adjusted and calibrated according to these customer-specific parameters.

3. To maintain the accuracy and quality of the flow values, we recommend re-calibrating the sensors every 24 months.

If your parameters have changed, contact em-tec-service@psgdover.com for a new adjustment!

4. If your parameters have not changed, you can carry out the calibration on site:

  • Use the same parameters as stated in the calibration report
  • Attach the sensor to the tube according to our recommendations
  • Perform the measurement for each of your chosen verification points and record the flow values
  • Generate a calibration report and check if the flow values are within the specific range

5. If measurements turn out to be outside the specifications, you can either:

  • Set a calibration factor to adapt the measure flow value or
  • Send the sensor to em-tec for readjustment

For more details, please refer to "BioProTT™ TechNote: How to Determine the Calibration Factor and How to Use It."